T2 measurement guide

Spin echo repeats Double measurements with 90° (phase 0°) → delay τ → 180° (phase 90°) → echo acquisition, estimating T2 from echo-amplitude decay.

Open T2 measurement ↗

Choose Japanese or English in the header. Settings and results are retained when switching between measurement pages.

Preparation and settings

Use values checked by frequency tuning and pulse calibration. Follow connection preparation and disconnect other apps using the same port first. All five measurement pages share a connection; switch after the current operation finishes without disconnecting.

SettingInitial valueCheck
τ start / end / step5 / 35 / 2 msKeep the analysis window within acquisition time
Frequency / 90° and 180°402000 Hz / 21 and 42 RF cyclesUse checked frequency and calibrated lengths
Repetition / scans200 ms / 64 per pointAllow sufficient recovery
FFT relative to echo center−4 to +4 msWindow must contain the center
Window / search bandHann / ±3 kHzCommon to all points

Defaults are starting examples. τ is an integer; up to 201 points, ending at the last on-grid delay. Repetition interval must exceed maximum τ+40.96 ms and allow sufficient sample recovery. Total/per-point estimates exclude communication overhead.

Operation

  1. Set frequency, calibrated pulse lengths, τ range, repetition interval and scans.
  2. Check the relative FFT window and search band, then click Start scan (走査開始).
  3. Inspect decay versus 2τ [ms]. Click a τ button to view its echo and FFT.
  4. Set the fit range in 2τ [ms] and click Update fit (近似を更新).
  5. If needed, change the reference point, or edit analysis settings and click Apply analysis (解析条件を適用).
  6. Save full JSON, T2 CSV and PNG as needed, then disconnect.

Stop after current measurement (現在の測定後に停止) waits for the active point. Errors do not retry and retain completed records. Save before starting another scan.

Read the results

S(2τ)=Ae−2τ/T2S(2\tau)=Ae^{-2\tau/T_2}

T2 is in ms; the screen shows T2, R², A, RMSE and point count. Background is fixed at zero. Both the axis and fit range use 2τ rather than τ. The axis starts at zero without adding a zero-time measurement.

A window that follows the echo

Waveform time zero is after the second pulse; the echo center is τ. For τ=15 ms and a −4 to +4 ms window, analysis uses 11–19 ms. Every window must contain the center and stay within 0–40.96 ms. With the default ±4 ms window, valid integer device delays are 4–36 ms.

A strong echo supplies a common frequency; all points are evaluated by complex magnitude at that frequency. 選択点を周波数基準にする selects it manually. Unlike T1, this uses no signed projection or common phase alignment.

Apply analysis (解析条件を適用) reanalyzes existing points without remeasuring. Echo waveforms show 0–40.96 ms and FFT spans ±10 kHz; the band marks the actual analysis window. The last 100 samples define the baseline, so inspect signal remaining in the tail for late echoes.

Fits are attempted from three distinct delays. Short ranges or poor agreement produce estimates with warnings; zero or flat data and other failed fits show a reason. See fit acceptance conditions.

Save and load

DownloadContents
Full JSON (全データJSON)All raw records, settings, reference frequency and fit range/results
T2 CSVτ, 2τ, echo amplitude, common frequency and actual analysis window
PNG保存Echo decay curve with an available fit
選択点の生データJSONOne selected record, readable in standard measurement

Open full JSON (全データJSONを開く) restores T2 scans up to 32 MiB and 201 points, including stopped scans. Without hardware, inspect waveforms/FFT, change fit ranges, reference frequency or analysis settings and save again. Results are recomputed from raw data rather than simply showing stored fits.

Import is disabled during measurement. Success replaces records and settings; failure keeps them. Zero-signal records can still be opened for waveform inspection. Use corresponding screens for other scan types such as T1. Save before closing; filenames use Japan time.

Try without hardware

T2 scan demo (T2走査デモ) uses synthetic echoes with T2=22 ms to try scanning, stopping, analysis, saving and loading. No device command is sent; screen and downloads identify DEMO data. This does not validate hardware accuracy.