T1 measurement guide

Inversion recovery repeats Double measurements with 180° (phase 0°) → delay τ → 90° (phase 0°) → FID acquisition, estimating T1 from signed signal recovery.

Open T1 measurement ↗

Use the language selector to switch between Japanese and English. Inputs, results, view, fit range and phase settings are retained when switching between the measurement pages.

Preparation and settings

Use values checked by frequency tuning and pulse calibration. Follow connection preparation, disconnecting other apps using the same port. Measurement, Frequency tuning, Pulse calibration, T1 measurement and T2 measurement share the connection; finish active operations before switching between them without disconnecting.

SettingInitial valueCheck
τ start / end / step5 / 100 / 5 msInclude negative-to-positive recovery
Frequency / 90° and 180°402000 Hz / 21 and 42 RF cyclesUse checked frequency and calibrated lengths
Repetition / scans200 ms / 64 per pointAllow sufficient recovery
FFT / window / search band1–5 ms / Hann / ±3 kHzCommon interval in FID after the second pulse

Defaults are starting examples. τ must be an integer from 1 to 32767 ms; at most 201 points, ending at the last on-grid delay. Repetition interval must exceed maximum τ+40.96 ms, and also allow sufficient sample recovery.

Total and per-point estimates update with the settings. They exclude communication overhead and are not demo playback durations.

Operation

  1. Set frequency, pulse lengths, τ range, repetition interval and scans.
  2. Check FFT and search-band settings, then click Start scan (走査開始).
  3. Inspect negative-to-positive recovery. Click a τ button to view its FID and FFT.
  4. Set the fit range in τ [ms] and click Update fit (近似を更新).
  5. If needed, review the phase reference/correction and click Reanalyze all points (全点を再解析).
  6. Save full JSON, T1 CSV and PNG as needed, then disconnect when finished, or switch to another measurement page without disconnecting.

Stop after current measurement (現在の測定後に停止) waits for the active point. Communication errors do not retry and retain completed records. Save before beginning a new scan.

Read the results

S(τ)=A−Be−τ/T1S(\tau)=A-Be^{-\tau/T_1}

T1 and τ are in ms. Inspect T1, R², A, B, B/A, RMSE and the point count. Ideal inversion gives B/A=2, but A and B are fitted independently. The axis begins at zero without adding a zero-delay point.

A common frequency and phase retain negative signals. For 選択点を位相基準にする, choose a sufficiently recovered positive FID. Extra phase correction applies to all points. Magnitude (絶対値) and Quadrature (直交成分) change only the display; the fit always uses signed signal.

Fitting requires at least five distinct delays, both signal signs and other acceptance conditions. A maximum τ below three times estimated T1 is a warning, not a rejection by itself. Coherence below 0.85, zero/flat signal or an inadequate fit produces an explanation. See the fitting API for exact conditions.

Save and load

DownloadContents
Full JSON (全データJSON)All raw records, measurement/analysis settings, phase reference and fit range/results
T1 CSVτ, signed signal, magnitude, quadrature, reference frequency and phase
PNGCurrent recovery view; an available fit is included in signed mode
選択点の生データJSONOne selected record for reanalysis in standard measurement

Open full JSON (全データJSONを開く) restores T1 scans up to 32 MiB and 201 points, including stopped scans. Without hardware, inspect FID/FFT, switch views, change phase references or fit ranges and save again. Fits are recomputed from raw records.

Import is disabled during measurement. Success replaces records and settings; failure keeps them. Zero-signal records can still be opened for waveform inspection. Use each corresponding screen for other scan types. Save before closing; filenames use Japan time.

Try without hardware

T1 scan demo (T1走査デモ) uses synthetic FID with T1=20 ms to try scanning, stopping, analysis and saving. No device command is sent; screen and downloads identify DEMO data. This does not validate hardware accuracy.