Skip to content

Pulse calibration guide

Keep frequency fixed and vary the first-pulse cycle count in Single measurements. Inspect Rabi oscillation peaks and troughs and use sine fits to estimate candidate 90° and 180° pulse lengths.

Open pulse calibration ↗

Switch between Japanese and English using the language selector. Inputs, results, signal view and phase settings remain available when switching between the measurement pages.

Preparation and settings

Enter the frequency checked by frequency tuning into Fixed frequency (固定周波数). Follow connection preparation and disconnect other apps using the same port. Measurement, Frequency tuning, Pulse calibration, T1 measurement and T2 measurement share the connection; finish active operations before switching, without disconnecting.

SettingInitial valueCheck
Start / end / step5 / 80 / 5 RF cyclesInclude the first peak and the region beyond it
Fixed frequency / first phase402000 Hz / 0°Use your checked frequency
Repetition / scans200 ms / 64 per pointAdjust for your sample
FFT / window / search band1–5 ms / Hann / ±3 kHzCommon to all points

The initial grid has 16 points. Up to 201 are allowed, ending at the last on-grid point. Total time estimates exclude communication overhead.

Operation

  1. Review frequency, scan range, scans and analysis settings; click Start scan (走査開始).
  2. Inspect peaks and troughs in the initial Magnitude view (絶対値).
  3. Switch to Signed signal (符号付き信号) to check the positive peak and subsequent sign reversal.
  4. Click a cycle count in the list to inspect its FID and FFT.
  5. When a fit is available, click 近似値を候補欄へ to display rounded integer 90°/180° candidates.
  6. Save and switch to Measurement without disconnecting, then enter your chosen values manually.

Copying candidates sends no command. It uses the fit for the currently selected view. The 21/42-cycle defaults in measurement screens are not established calibration values.

Read the results

Magnitude folds negative signals upward. Do not identify 90°/180° solely from its peaks and troughs; inspect signed data and phase agreement too. The axis starts at zero without adding a zero-cycle measurement.

S(p)=Asin⁡ ⁣(πp2p90)+B,p180=2p90S(p)=A\sin\!\left(\frac{\pi p}{2p_{90}}\right)+B,\qquad p_{180}=2p_{90}

p is the first-pulse RF cycle count. Magnitude fitting uses the absolute sine. Fits are attempted from three distinct lengths; phase inconsistency, low agreement and extrapolation generate warnings. Inspect the curve, R² and individual waveforms.

Measured candidates use the first positive peak and the next positive-to-negative crossing, interpolated between neighboring points. They are distinct from sine-fit candidates. If no estimate is possible, such as for zero signal, the reason is shown.

Common phase reference

All points use one frequency and phase reference, retaining sign reversals. Select a point and click 選択点を位相基準にする, or enter an extra phase correction and click 全点を再解析. 自動基準に戻す restores automatic selection. Changing the reference clears candidate fields; check and copy them again. This does not remeasure or alter raw samples.

Save and load

DownloadContents
Full JSON (全データJSON)All raw records, settings, phase reference, signed analysis and fits
Calibration CSV (校正CSV)Cycles, signed signal, magnitude, quadrature, reference frequency/phase
PNGCurrent view with an available sin or |sin| fit
選択点の生データJSONOne selected record, readable in standard measurement

Open full JSON (全データJSONを開く) restores a pulse scan up to 32 MiB and 201 points, including stopped scans. Without hardware, switch views, inspect individual waveforms, change the phase reference/correction and save again. Analysis is recomputed from saved raw data.

Import is disabled during scanning. Success replaces records and settings; invalid files retain them. Zero-signal records can still be opened for waveform inspection without a reference or fit. Open other archive types in their corresponding screens.

Stopping and trying without hardware

Stop after current measurement (現在の測定後に停止) waits for the active point. Errors do not retry and retain completed records. Save before another scan or closing the page. Filenames use Japan time.

Pulse scan demo (パルス走査デモ) uses a synthetic signal with 90°=21 cycles. It sends no device commands and marks results as simulated. It does not validate physical calibration accuracy.

T1 guide · T2 guide · Standard measurement guide